tina51e1-c (2011-12) 21
4 Operation
Put the gauge into operation. If you are using an INFICON con-
troller, define the measurement range ( [1, 2, 3]).
A warm-up time of at least ½ hour should be allowed; for precise
pressure measurements a warm-up time of at least 2 hours is
required.
If the gauge is used for fast downstream pressure con-
trol we recommend setting its signal filter to "fast".
The filter can be set via the RS232C interface or the
diagnostic port ( [4]).
4.1 Status Indication
S
P
1
S
P
2
S
T
A
T
U
S
SP ZERO
DIAG
I/O
LED LED status Meaning
<STATUS> off no supply voltage
lit solid green measurement mode
blinking green
short blinks
long blinks
warning, over/underrange
warming up
lit solid red error
<SP1> lit green green
p
setpoint 1
blinking green waiting for setpoint 1 input
off p > setpoint 1
<SP2> lit solid green
p
setpoint 2
blinking green waiting for setpoint 2 input
off p > setpoint 2